Newer Generation DePuy INHANCE™ Modular Anatomic Convertible Polyethylene Glenoid and Platform Baseplate have Improved Resistance to Micromotion and Dissociation at the Polyethylene-Metal Junction.

Document Type

Article

Publication Date

11-10-2026

Publication Title

Journal of shoulder and elbow surgery / American Shoulder and Elbow Surgeons ... [et al.]

Abstract

INTRODUCTION: Metal-backed glenoid (MBG) designs in anatomic total shoulder arthroplasty (aTSA) were introduced in the 1980s to address concerns with the high rate of loosening of all-polyethylene glenoid components, offering potential advantages of durable glenoid fixation and easy revisability, but have historically demonstrated concerning rates of mechanical failure including polyethylene (PE) liner dissociation. Recently, platform baseplates have been introduced which can be used in anatomic or reverse TSA configurations. Platform baseplates may improve clinical performance, but more biomechanical analysis is needed to support their implant design improvements over the failed MBG designs of the past. The purpose of this study was to biomechanically evaluate a newer generation DePuy INHANCE™ modular anatomic convertible glenoid PE liner and platform baseplate and its ability to withstand mechanical forces and resist polyethylene dissociation when subjected to worst-case loading conditions.

METHODS: The DePuy INHANCE™ convertible glenoid + platform baseplate were tested for resistance to micromotion from repetitive loading and compared to an all-polyethylene glenoid component. The study implant was also evaluated for its ability to withstand tensile displacement and polyethylene dissociation from shear and axial forces and compared against another commonly used modular anatomic glenoid PE liner and platform baseplate implant. Micromotion testing was performed in accordance with the standardized modular dissociation test method (ASTM F028-17) that was developed specifically for modular anatomic glenoid systems.

RESULTS: The average tensile displacement of the INHANCE™ study implant after cyclic rocking was significantly less than the all-polyethylene glenoid component (0.1 mm vs. 0.4 mm; p=0.03). The INHANCE™ polyethylene-metal junction had a significantly greater resistance to modular dissociation under shear loading (1104N, 95% CI [1068, 1140] vs. 713N, 95% CI [687, 739]; p< 0.001) and axial loading (512N, 95% CI [493, 531] vs. 250, 95% CI [205, 295]; p< 0.001) than was demonstrated by a competitive modular anatomic glenoid PE liner and platform baseplate system.

CONCLUSION: Our investigation shows that the DePuy INHANCE™ modular anatomic convertible glenoid PE liner and platform baseplate demonstrates favorable biomechanical performance against micromotion and polyethylene dissociation. These results should be considered when evaluating the current and future role of platform baseplates in aTSA.